Characterization and calibration of a multilayer coated Wolter optic for an imager on the Z-machine at Sandia National Laboratories.

M WuB KozioziemskiJ K VogelP LakeJ R FeinDavid J AmplefordC J BourdonJ AyersP BellD BradleyC C WaltonLouisa PickworthM J PivovaroffAndrew AmesR BruniS RomaineK KilaruB Ramsey
Published in: The Review of scientific instruments (2018)
The need for a time-resolved monochromatic x-ray imaging diagnostic at photon energies >15 keV has motivated the development of a Wolter optic to study x-ray sources on the Z-machine at Sandia National Laboratories. The work is performed in both the LLNL's x-ray calibration facility and SNL's micro-focus x-ray lab. Characterizations and calibrations include alignment, measurement of throughput within the field of view (FOV), the point-spread function within the FOV both in and out of focus, and bandpass in the FOV. These results are compared with ray tracing models, showing reasonable agreement.