Evaluation of a Chair-Mounted Passive Trunk Orthosis: A Pilot Study on Able-Bodied Subjects.

Ahmad Zahid RaoMuhammad Abul Hasan
Published in: Sensors (Basel, Switzerland) (2021)
Trunk stability is important for adequate arm function due to their kinematic linkage. People with Duchenne muscular dystrophy (DMD) can benefit from trunk-assistive devices for seated daily activities, but existing devices limit trunk movement to forward bending. We developed a new trunk orthosis that has spring and pulley design. This study evaluated orthosis performance with 40 able-bodied subjects under with and without orthosis condition in 20 seated tasks for trunk rotation, forward bending, and side bending movements. Subjects adopted static posture in specific trunk orientation while their muscle activity was recorded. They also rated the subjective scales of perceived exertion and usability. A percent change in muscle activity for each task, due to orthosis use, is reported. Significant muscle activity reductions up to 31% and 65% were observed in lumbar and thoracic erector spinae muscles, respectively. Using three-way ANOVA, we found these reductions to be specific to the task direction and the choice of upper limb that is used to perform the asymmetric tasks. A total of 70% participants reported acceptable usability and ~1-point increase in exertion was found for maximum voluntary reaching with the orthosis. The outcomes of this study are promising, though tested on able-bodied subjects. Hence, orthosis mounted on wheelchairs should be further evaluated on DMD patients.