Montel mirror based collimating analyzer system for high-pressure resonant inelastic X-ray scattering experiments.
Jin Kwang KimDiego CasaXianrong HuangThomas GogB J KimJung Ho KimPublished in: Journal of synchrotron radiation (2020)
Resonant inelastic X-ray scattering (RIXS) is increasingly playing a significant role in studying highly correlated systems, especially since it was proven capable of measuring low-energy magnetic excitations. However, despite high expectations for experimental evidence of novel magnetic phases at high pressure, unequivocal low-energy spectral signatures remain obscured by extrinsic scattering from material surrounding the sample in a diamond anvil cell (DAC): pressure media, Be gasket and the diamond anvils themselves. A scattered X-ray collimation based medium-energy resolution (∼100 meV) analyzer system for a RIXS spectrometer at the Ir L3-absorption edge has been designed and built to remediate these difficulties. Due to the confocal nature of the analyzer system, the majority of extrinsic scattering is rejected, yielding a clean low-energy excitation spectrum of an iridate Sr2IrO4 sample in a DAC cell. Furthermore, the energy resolution of different configurations of the collimating and analyzing optics are discussed.