Point-of-Care Versus Central Laboratory Measurements of Hemoglobin, Hematocrit, Glucose, Bicarbonate and Electrolytes: A Prospective Observational Study in Critically Ill Patients.
Jérôme Allardet-ServentMelissa LebsirChristian DubrocaMartine FabrigouleSylvie JordanaThomas SignouretMatthias CastanierGuillemette ThomasRettinavelou SoundaravelouAnne LepidiLaurence DelapierreGuillaume PenarandaPhilippe HalfonJean-Marie SeghboyanPublished in: PloS one (2017)
Compared to central Laboratory analyzers, the POC Siemens RAPIDPoint 500 blood gas system satisfied the CLIA criteria of interchangeability for all tested parameters, except for hemoglobin. These results are warranted for our own procedures and devices. Bearing these restrictions, we recommend clinicians to initiate an appropriate therapy based on POC testing without awaiting a control measurement.