Magnitude dependent discordance in liver stiffness measurements using elastography point quantification with transient elastography as the reference test.
Ivica GrgurevićNermin SalkicTonci BozinSanda MustapicVladimir MaticIvo Dumic-CuleIda Tjesic DrinkovicTomislav BokunPublished in: European radiology (2018)
• ElastPQ appears to be reliable method for assessment of liver fibrosis, with data presented here mostly applicable to NAFLD. • LSM values produced by TE and ElastPQ are NOT interchangeable-in values < 10 kPa, they are similar, but in values > 10 kPa, they appear to be increasingly and significantly different. • Diagnostic performance of ElastPQ for sub-classification of patients with compensated advanced chronic liver disease should be furtherly investigated.