Login / Signup

Biofunctionalization of surfaces to minimize undesirable effects in cardiovascular assistance devices.

Rosa Correa Leoncio de SaAron Jose Pazin de AndradeVagner Roberto AntunesCecilia SalvadoriFernanda de Sa TeixeiraEvaldo Jose CoratJoao Roberto MoroEduardo Guy Perpetuo BockVladimir Jesus Trava-Airoldi
Published in: Artificial organs (2023)
After twelve weeks, the implants were extracted and subjected to qualitative analysis by Scanning Electron Microscopy under low vacuum and X-ray Energy Dispersion. The regions that remained in contact with the wall of the aorta showed encapsulation of the endothelial tissue. TSB implants, although superhydrophilic, have proven that the DLC coating inhibits the adhesion of biological material, prevents abrasive wear and delamination, as observed in the TS and TSA implants. Pseudo- neointimal layers were heterogeneously identified in higher concentration on Test Surfaces.
Keyphrases