The effect of prolonged laser activation on irrigation fluid temperature: an in vitro experimental study.
Tsaturyan ArmanAngelis PeteinarisLampros PantazisErgina FarsariKonstantinos PagonisConstantinos AdamouAthanasios VagionisAnastasios NatsosEvangelos LiatsikosPanagiotis KallidonisPublished in: World journal of urology (2022)
Our study found that continuous laser activation with as less as 12 W using 10 ml/min outflow rate increased the irrigation fluid temperature above the threshold only after 1 min. In the current experimental setup, with the fluid outflow rate of 20 and 30 ml/min safe laser activation with 60 W and 40 W (temperature < 43 °C) can be achieved reaching the same total emitted energy as with 12 W in significantly shorter time period.