Total Knee Arthroplasty in Haemophilia: Long-Term Results and Survival Rate of a Modern Knee Implant with an Oxidized Zirconium Femoral Component.
Christian CarulliMatteo InnocentiRinaldo TambascoAlessandro PerroneCivinini RobertoPublished in: Journal of clinical medicine (2023)
(1) Background: Total Knee Arthroplasty (TKA) in patient with haemophilia (PWH) has usually been performed with the use of cobalt-chrome femoral and titanium tibial components, coupled with standard polyethylene (PE) inserts. The aim of this retrospective study was to evaluate the long-term outcomes and survival rates of TKA in a series of consecutive PWH affected by severe knee arthropathy at a single institution. (2) Methods: We followed 65 patients undergoing 91 TKA, implanted using the same implant, characterized by an oxidized zirconium femoral component, coupled with a titanium tibial component, and a highly crosslinked PE. At 1, 6, and 12 months; then every year for 5 years; and finally, every other 3 years, all patients were scored for pain (VAS), function (HJHS; KSS), ROM, and radiographic changes. Kaplan-Meier survivorship curves were used to calculate the implant survival rates. (3) Results: The mean follow-up was 12.3 years (4.2-20.6). All clinical and functional scores improved significantly from preoperatively to the latest follow-up (VAS: from 6.9 to 1.3; HJHS: from 13.4 to 1.9; KSS: from 19.4 to 79; ROM: from 42.4° to 83.6°). The overall survivorship of the implants was 97.5% at the latest follow-up. (4) Conclusions: The present series showed a high survival rate of specific implants potentially linked to the choice of an oxidized zirconium coupled with a highly crosslinked PE. We promote the use of modern implants in these patients in order to ensure long-lasting positive outcomes.