Hippocampal Sclerosis Detection with NeuroQuant Compared with Neuroradiologists.
Shreya LouisMárcia E MoritaStephen E JonesD VeghWilliam BingamanIngmar BluemckeNancy A ObuchowskiFernando CendesLara E JehiPublished in: AJNR. American journal of neuroradiology (2020)
Visual MR imaging analysis by a neuroradiologist with expertise in epilepsy had a higher sensitivity than did NeuroQuant analysis, likely due to the inability of NeuroQuant to evaluate changes in hippocampal T2 signal or architecture. Given that there was no significant difference in specificity between NeuroQuant analysis and visual MR imaging analysis, NeuroQuant can be a valuable tool when the results are positive, particularly in centers that lack neuroradiologists with expertise in epilepsy, to help identify and refer candidates for temporal lobe epilepsy resection. In contrast, a negative test could justify a case referral for further evaluation to ensure that false-negatives are detected.