Login / Signup

Direct measurement and correction of both megavoltage and kilovoltage scattered x-rays for orthogonal kilovoltage imaging subsystems with dual flat panel detectors.

Hiraku IraminaMitsuhiro NakamuraTakashi Mizowaki
Published in: Journal of applied clinical medical physics (2020)
MV- and kV-scatter could be corrected by subtracting the scatter maps from the projections, and the correction improved the intensity ratios of the GMs.
Keyphrases
  • monte carlo
  • high resolution
  • high intensity
  • image quality
  • magnetic resonance