Accuracy Assessment of Oura Ring Nocturnal Heart Rate and Heart Rate Variability in Comparison With Electrocardiography in Time and Frequency Domains: Comprehensive Analysis.
Rui CaoIman AzimiFatemeh SarhaddiHannakaisa Niela-VilenAnna M AxelinPasi LiljebergAmir M RahmaniPublished in: Journal of medical Internet research (2022)
The Oura Ring could accurately measure nocturnal HR and RMSSD in both the 5-minute and average-per-night tests. It provided acceptable nocturnal AVNN, pNN50, HF, and SDNN accuracy in the average-per-night test but not in the 5-minute test. In contrast, the LF and LF:HF ratio of the ring had high error rates in both tests.