Detection of direct oral anticoagulants with the diluted Russel's viper venom time.
Tristan KlöterMichael MetzeRonny KunzeStephan StöbeThomas SiegemundAnnelie SiegemundReinhard HenschlerUlrich LaufsSirak PetrosChristian PfrepperPublished in: International journal of laboratory hematology (2024)
The dRVVT confirm assay (LA2) reliably detects residual DOAC plasma levels ≥30 ng/mL and could be useful to rapidly rule out relevant DOAC activity in emergency situations and to guide treatment decisions.