A calibration framework for the determination of accurate collision cross sections of polyanions using polyoxometalate standards.
Sébastien HupinHélène LavanantSéverine RenaudineauAnna ProustGuillaume IzzetMichael GroesslCarlos AfonsoPublished in: Rapid communications in mass spectrometry : RCM (2018)
Our novel calibration strategy for determination of CCS values of multiply negatively charged ions on TWIM-MS devices based on DT CCS (N2 ) of standard POM structures covered a wider range of CCS and improved the accuracy to 2.1% relative error on average compared with 6.9% using polyalanine and dextran calibration.