Comparative evaluation of the diagnostic accuracies of four different malaria rapid diagnostic test kits available in Ghana.

Enoch AninagyeiJohn Gameli DekuKeren Trishia YemofioEkua QuainooKofi Adjei NtiriEvelyn YaroPriscilla EssandohHubert Kwame AgbogliRichard Harry Asmah
Published in: PloS one (2024)
Malaria rapid diagnostic test (mRDT) kit is one of the techniques for diagnosing malaria. Due to its inherent advantages over the microscopy technique, several brands of the kit have flooded malaria endemic countries, without prior in-country evaluation. Two of such mRDT kits are Oscar (India) and Standard Q (Korea Republic). In this study, the performance of Oscar and Standard Q mRDT kits were compared to First Response (India) and CareStart (USA) mRDTs, which have been evaluated and deployed for use approved by the Ministry of Health (MOH). In this comparative study, whole blood samples were collected from patients suspected of malaria. Plasmodium falciparum was detected in each sample using nested polymerase chain reaction (nPCR), microscopy and the four mRDTs. The sensitivities, specificities, accuracies, positive and negative predictive values and accuracies of the mRDTs were determined using nPCR as a reference technique. Kappa statistic was used to determine the level of agreement among the techniques. Two hundred (200) blood samples were analyzed in this study. The overall detection rates of P. falciparum by microscopy, First Response, CareStart, Oscar-PfHRP2, Standard Q mRDT kits and nPCR were 31.5%, 34.5%, 33.5%, 32%, 31% and 43% (x2 = 6.1, p = 0.046), respectively. The accuracies of CareStart and First Response were comparable (90.5% vs. 89.5%). Further, comparing their sensitivities, Oscar-PfHRP2 was 74.4% (95% confidence interval (CI): 63.9-83.2) while that of Standard Q was 72.1% (95% CI: 61.4-81.2), with comparable accuracies (Oscar-PfHRP2-89% and Standard Q -88%). Apart from First Response that was 98.3% specific, the others were 100% specific. Kappa test revealed perfect diagnostic agreement (κ = 0.90-0.98) among the four mRDTs. That notwithstanding, Oscar-PfHRP2 agreed better with CareStart (κ = 0.94) and First Response (κ = 0.92) compared to the agreement between Standard Q and, CareStart (κ = 0.92) and First Response (κ = 0.90). Taken together, the diagnostic performance of the four mRDT kits were statistically similar. That notwithstanding, new mRDT kits should be evaluated prior to deployment for use.
Keyphrases