Inflated Gamma Interferon Response with QuantiFERON-TB Gold Plus Using the Automated Liaison XL Analyzer: a Testing Algorithm To Mitigate False-Positive Results in Low-Incidence Settings.
Vladimir BuronNiaz BanaeiPublished in: Journal of clinical microbiology (2023)
The Liaison XL chemiluminescence immunoassay (CLIA) analyzer allows total automation of gamma interferon (IFN-γ) measurement for the QuantiFERON-TB Gold Plus assay (QFT-Plus) that is used to diagnose Mycobacterium tuberculosis infection. To evaluate CLIA accuracy, plasma samples from 278 patients undergoing QFT-Plus testing were first tested with an enzyme-linked immunosorbent assay (ELISA; 150 negatives and 128 positives) and subsequently with the CLIA. Three strategies to mitigate false-positive CLIA results were investigated in 220 samples with borderline-negative ELISA results (TB1 and/or TB2, 0.1 to 0.34 IU/mL). The Bland-Altman plot of difference versus average of the two IFN-γ measurements in the Nil and antigen (TB1 and TB2) tubes showed higher IFN-γ measurements across the range of values with the CLIA than with the ELISA. Bias was 0.21 IU/mL (standard deviation, 0.61; 95% confidence interval [CI], -1.0 to 1.41). Linear regression of difference versus average had a slope of 0.08 (95% CI, 0.05 to 0.10), which was significantly nonzero ( P < 0.0001). The CLIA had positive and negative percent agreement levels with the ELISA of 91.7% (121/132) and 95.2% (139/146), respectively. In borderline-negative samples tested with ELISA, CLIA was positive in 42.7% (94/220). CLIA with a standard curve resulted in 36.4% (80/220) positivity. Retesting CLIA false positives (TB1 or TB2 range, 0 to ≤1.3 IU/mL) with ELISA reduced false positives by 84.3% (59/70). Retesting with CLIA reduced the false-positive rate by 10.4% (8/77). Implementing the Liaison CLIA for QFT-Plus in low-incidence settings risks falsely elevating conversion rates and overburdening clinics and overtreating patients. Confirming borderline positives with ELISA is a viable strategy to mitigate false-positive CLIA results.
Keyphrases