A Novel Control Method for Rotary Blood Pumps as Left Ventricular Assist Device Utilizing Aortic Valve State Detection.

Dmitry PetukhovLeonie KornMarian WalterDmitry V Telyshev
Published in: BioMed research international (2019)
A novel control method for rotary blood pumps is proposed relying on two different objectives: regulation of pump flow in accordance with desired value and the maintenance of partial support with an open aortic valve by the variation of pump speed. The estimation of pump flow and detection of aortic valve state was performed with mathematical models describing the first- and second generation of Sputnik rotary blood pumps. The control method was validated using a cardiovascular system model. The state of the aortic valve was detected with a mean accuracy of 91% for Sputnik 1 and 96.2% for Sputnik 2 when contractility, heart rate, and systemic vascular resistance was changed. In silico results for both pumps showed that the proposed control method can achieve the desired pump flow level and maintain the open state of the aortic valve by periodically switching between two objectives under contractility, heart rate, and systemic vascular resistance changes. The proposed method showed its potential for safe operation without adverse events and for the improvement of chances for myocardial recovery.