Measurements of Plasma-Free Metanephrines by Immunoassay Versus Urinary Metanephrines and Catecholamines by Liquid Chromatography with Amperometric Detection for the Diagnosis of Pheochromocytoma/Paraganglioma.

Wolfgang RaberHans KotalRodrig MarculescuChristian ScheubaMartin B NiederleAlexandra Kautzky-WillerMichael R Krebs
Published in: Journal of clinical medicine (2020)
Studies conflict concerning the use of enzyme immunoassays (EIA) for plasma free metanephrines (P-MNs) vs. other methods for pheochromocytoma/paraganglioma (PPGL) diagnosis. We compared commercially available EIAs for P-MNs with high-pressure liquid chromatography (HPLC) for 24 h-urinary MNs (U-MNs) and -catecholamines (U-CATs). 943 (565 female, 378 male) patients (54 PPGL, 889 Non-PPGL) were studied. Simultaneous measurements of all parameters analyzed at the central lab of our university hospital was mandatory for inclusion. Sensitivity of P-MNs (94.4%) was similar to that of U-MNs (100%), and both were higher than of U-CATs (77.8%), specificity of P-MNs (100%) higher than of U-MNs (73.6%), and similar to U-CATs (99.8%). With the recently proposed downward adjusted ULN of P-MNs to correct for the reported negative bias of the EIAs sensitivity (98.1%) raised non-significantly, but specificity decreased significantly (94.8%). Areas under receiver-operating characteristic curves indicated comparable diagnostic performance of P-MNs (0.989) vs. U-MNs (0.995), both better than U-CATs (0.956). In summary, the EIAs to measure P-MNs performed similarly to U-MNs by HPLC, and both better than U-CATs by HPLC. The post-test probability of PPGL given a positive test result was best for P-MNs, and higher than for the other pairs of analytes. Downward corrections of ULN of P-MNs did not improve test performances.
Keyphrases