Efficient pulse sequence design framework for high-dimensional MR fingerprinting scans using systematic error index.

Siyuan HuZhilang QiuRichard James AdamsWalter ZhaoRasim BoyaciogluDaniela CalvettiDebra F McGivneyDan Ma
Published in: Magnetic resonance in medicine (2024)
We developed an efficient method for estimating real-world errors in MRF scans with high computational efficiency. Our results illustrate that the SEI model could approximate errors both qualitatively and quantitatively. We also proved the practicality of the SEI model of optimizing sequences for high-dimensional MRF frameworks with manageable computational power. The optimized high-dimensional MRF scans exhibited enhanced robustness against undersampling and system imperfections with faster scan times.
Keyphrases