Method development for quantification of quizartinib in rat plasma by liquid chromatography/tandem mass spectrometry for pharmacokinetic application.

Essam A AliMuzaffar IqbalGamal A E MostafaKhalid A Al-RashoodToqa El-Nahhas
Published in: Biomedical chromatography : BMC (2017)
Quizartinib is a highly potent inhibitor of the fms-like tyrosine kinase receptor, which is one of the most commonly mutated genes in acute myeloid leukemia. Quizartinib has shown a significant antileukemic clinical influence among relapsed/refractory acute myeloid leukemia patients. This study aimed at developing and validating an analytical method for the measurement of quizartinib in rat plasma using liquid chromatography-tandem mass spectrometry (LC-MS/MS). The method was validated according to US Food and Drug Administration guidelines, and the results obtained in this work met the set criteria. Liquid-liquid extraction was used and chromatographic separation was achieved on a BEHTM C18 column. Detection of quizartinib was achieved in multiple reaction monitoring mode using positive-ion mode electrospray ionization. The MS/MS ion transitions at mass-to-charge ratios (m/z) of 561.129/114.09 and 441.16/84.03 were monitored for quizartinib and ibrutinib, respectively. The linear detection range was 2-1000 ng/mL (r > 0.998), with intra- and inter-day assay precisions ≤13.07 and 13.17%, respectively. This rapid, simple and sensitive method was validated and successfully applied to the pharmacokinetic study of quizartinib in rat samples.