Femoral Head Penetration Rates and Oxidation of Highly Cross-Linked Polyethylene Hip Liners Implanted More Than 10 Years.

Tabitha DerrDaniel W MacDonaldArthur L MalkaniGregg R KleinSteven M Kurtz
Published in: The Journal of arthroplasty (2024)
Oxidation was found to be higher in the annealed HXLPE; however, this does not seem to be associated with greater wear as we found the average penetration rates for the cohorts were low, and the penetration rates were similar between the annealed and remelted cohorts.