Login / Signup

In Silico Modelling to Assess the Electrical and Thermal Disturbance Provoked by a Metal Intracoronary Stent during Epicardial Pulsed Electric Field Ablation.

Ana Gonzalez-SuarezJuan J PérezBarry O'BrienMuhammad Adnan Elahi
Published in: Journal of cardiovascular development and disease (2022)
The presence of a metal intracoronary stent near the ablation device during PEF ablation simply 'amplifies' the E-field distortion already caused by the presence of the vessel. This distortion may involve moderate heating (<48 °C) in the tissue between the artery and ablation electrode without associated thermal damage.