Increasing the Upper Mass/Charge Limit of a Quadrupole Ion Trap for Ion/Ion Reaction Product Analysis via Waveform Switching.

Kenneth W LeeGregory S EakinsMark S CarlsenScott A McLuckey
Published in: Journal of the American Society for Mass Spectrometry (2019)
Quadrupole ion traps (QITs) are versatile platforms for performing experiments with gas-phase ions due to their abilities to store ions of both polarities and to conduct MSn experiments. The QIT is particularly useful as a reaction cell for ion/ion reactions. In the case of an ion/ion reaction experiment in a QIT, multiply charged reactant ions may initially be of relatively low m/z (e.g., m/z < 1000) whereas the product ions can be one or more orders of magnitude higher in m/z (e.g., m/z > 100,000). Several factors can limit the m/z range over which an ion/ion reaction experiment can be conducted. These include (1) the efficiency of the detector, (2) the m/z range over which oppositely charged ions can be mutually stored, and (3) the m/z range over which ions can be mass selectively ejected into an external detector. High-frequency waveforms provide larger m/z trapping ranges for mutual storage of oppositely charged ions whereas low-frequency waveforms provide better trapping for very high m/z product ions. Presented here is a method that switches from a high-frequency sine wave prior to and during an ion/ion reaction to a low-frequency square wave to eject low m/z reagent ions and improves confinement of the product ions before mass-selective ejection by scanning the frequency of the square wave. This approach addresses the third issue, which is the primary limiting factor with QITs operated at high RF (e.g., > 900 MHz). Graphical Abstract.
Keyphrases
  • high frequency
  • quantum dots
  • aqueous solution
  • mass spectrometry
  • transcranial magnetic stimulation
  • magnetic resonance imaging
  • stem cells
  • liquid chromatography
  • bone marrow