Tear meniscus height agreement and reproducibility between two corneal topographers and spectral-domain optical coherence tomography.

Ivo SoaresEva RamalhoFrancisco Miguel BrardoAmélia Fernandes Nunes
Published in: Clinical & experimental optometry (2024)
The Oculus Keratograph 5 M, Medmont Meridia, and Spectralis SD-OCT provide reliable and reproducible inter-observer TMH measurements. Inter-device reliability is moderate, with a close correlation between Spectralis SD-OCT and the Oculus Keratograph 5 M. Oculus Keratograph 5 M and Medmont Meridia are repeatable devices appropriate for the measurement of TMH, but they are not interchangeable in clinical practice.