Interobserver Agreement and Correlation of an Automated Algorithm for B-Line Identification and Quantification With Expert Sonologist Review in a Handheld Ultrasound Device.
Christopher L MooreJing WangAnthony J BattistiAlvin ChenJonathan FinckeAnita WangMichael WagnerBalasundar RajuCristiana BaloescuPublished in: Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine (2021)
This study demonstrates a high correlation between point-of-care ultrasound experts and an automated algorithm to identify and quantify B-lines using an HHUD. Future research may incorporate this HHUD in clinical studies in multiple settings and users of varying experience levels.