A biopsy-needle compatible varifocal multiphoton rigid probe for depth-resolved optical biopsy.
Ang LiGunnsteinn HallDefu ChenWenxuan LiangBo NingHonghua GuanXingde LiPublished in: Journal of biophotonics (2018)
In this work, we report a biopsy-needle compatible rigid probe, capable of performing three-dimensional (3D) two-photon optical biopsy. The probe has a small outer diameter of 1.75 mm and fits inside a gauge-14 biopsy needle to reach internal organs. A carefully designed focus scanning mechanism has been implemented in the rigid probe, which, along with a rapid two-dimensional MEMS scanner, enables 3D imaging. Fast image acquisition up to 10 frames per second is possible, dramatically reducing motion artifacts during in vivo imaging. Equipped with a high-numerical aperture micro-objective, the miniature rigid probe offers a high two-photon resolution (0.833 × 6.11 μm, lateral × axial), a lateral field of view of 120 μm, and an axial focus tuning range of 200 μm. In addition to imaging of mouse internal organs and subcutaneous tumor in vivo, first-of-its-kind depth-resolved two-photon optical biopsy of an internal organ has been successfully demonstrated on mouse kidney in vivo and in situ.