The Length Versus Diameter Relationship and Radial Force Properties of the Amplatzer™ Vascular Plug Type IV: Observations for Oversizing.

Jenny E ZablahRachel FryDerek EilersGareth MorganNeil WilsonMichael Ross
Published in: Cardiovascular engineering and technology (2019)
The AVP IV device has a reliable length for diameter relationship. A counterintuitive property of the AVP IV with regards to radial force for device size was found. The 5 mm AVPIV was found to exert the highest radial force and stiffness compared to the other devices. The 7 and 8 mm AVPIV devices were consistently found to exert less radial force. This property suggests that oversizing an AVP IV could be safe and effective, potentially decreasing total devices used, cost, and overall procedure time.