Login / Signup

Finite element analysis comparison between superior clavicle locking plate with and without screw holes above fracture zone in midshaft clavicular fracture.

Nachapan PengrungNatthaphop LakdeeChedtha PuncreobutrBoonrat LohwongwatanaPaphon Sa-ngasoongsong
Published in: BMC musculoskeletal disorders (2019)
The weakest link of the SCLP construct in comminuted midshaft clavicular fracture fixation is the SHFZ, especially in the cantilever bending load. Additionally, the biomechanical property of the SCLP without SHFZ model (No-SHFZ plate) is superior to the standard SCLP model (SHFZ plate), with a significantly lower peak stress on the SHFZ location in all loading conditions. We recommend a new SCLP design with SHFZ to prevent implant failure and improve surgical outcomes.