Login / Signup

Comparison of cardiac function index derived from femoral and jugular indicator injection for transpulmonary thermodilution with the PiCCO-device: A prospective observational study.

Alexander HernerMarkus HeilmaierUlrich MayrRoland M SchmidWolfgang Huber
Published in: PloS one (2018)
While the last PiCCO algorithm obviously corrects GEDVI for femoral indicator injection, this correction is not applied to CFI. Therefore, femoral TPTD indicator injection results in substantially lower values for CFI compared to TPTD using a jugular CVC. Necessarily, uncorrected CFI-values derived from femoral TPTD are misleading and have to be corrected.
Keyphrases
  • ultrasound guided
  • machine learning