LOCAL DIAGNOSTIC REFERENCE LEVELS FOR PEDIATRIC RETROGRADE WEDGE PORTOGRAPHY INTERVENTIONAL PROCEDURES USING A DOSE MONITORING SOFTWARE AT A TRANSPLANTATION INSTITUTE.

Roberta GerasiaC CannataciG S GalloC TafaroL MaruzzelliK CortisR Miraglia
Published in: Radiation protection dosimetry (2022)
The aim of this work was to establish local diagnostic reference levels (DRLs) for retrograde wedge portography (RWP) performed on pediatric patients assessing the usefulness of radiation dose monitoring software in the establishing process. Between September 2016 and April 2020, 66 consecutive RWP were performed at a transplantation institute and were included in our study. Patients were divided in three groups according to age: n = 25 infants, n = 20 middle childhood and n = 21 early adolescence. The third quartile of both Air Kerma at the reference point (Ka,r) and air kerma-area product (PKA) were evaluated to establish local DRLs (lDRLs). In addition, to control high Ka,r levels during procedures, the software notified to operators if Ka,r exceeded the dose 'alert' threshold set at 2 Gy. lDRLs were established for all three groups using PKA and Ka,r: infant group: 5.6 Gy.cm 2 and 0.034 Gy; middle childhood: 6.4 Gy.cm2 and 0.018 Gy and early adolescence: 12.8 Gy.cm2 and 0.059 Gy. The dose threshold 'alert' was never encountered (alert quota: 0%). The dose monitoring system supports the feasibility of accurate and easier lDRLs' establishment.