Login / Signup

Transferability of patients for radiation treatment between unmatched machines.

Joseph J FoySerpil K DoganPoonam YadavBharat B MittalIndra J Das
Published in: Journal of applied clinical medical physics (2022)
Despite small differences in calculated dose, transferring patients between two unmatched Elekta machines with similar multileaf collimator (MLC)-head for target coverage and minimum changes in OAR dose is possible for a limited number of fractions (≤3) to improve clinical flexibility and institutional throughput along with patient satisfaction. A similar study could be carried out for other machines for operational throughput.
Keyphrases
  • end stage renal disease
  • ejection fraction
  • newly diagnosed
  • chronic kidney disease
  • patient satisfaction
  • prognostic factors
  • peritoneal dialysis
  • healthcare
  • radiation therapy
  • health insurance
  • optic nerve