Development of a large diameter in vitro flow loop thrombogenicity test system.
Carlos SernaAnna ParrishMehulkumar PatelKeerthana SrinivasanRichard MalinauskasQijin LuMegan A JamiolkowskiPublished in: Artificial organs (2024)
The 9.5 -mm ID test system can effectively differentiate materials of varying thrombogenic potentials when appropriate pump occlusion settings and donor-specific anticoagulation are used. This system is being assessed in an interlaboratory study to develop standardized best practices for performing in vitro dynamic thrombogenicity testing of medical devices and materials.