Login / Signup

NEMA NU2-2012 performance measurements of the United Imaging uPMR790: an integrated PET/MR system.

Shuguang ChenYushen GuHaojun YuXin ChenTuoyu CaoLingzhi HuHongcheng Shi
Published in: European journal of nuclear medicine and molecular imaging (2021)
The PET performance of the newly introduced integrated PET/MR system is not significantly affected by the simultaneous operation of an MR sequence (2-point DIXON sequence). Measurement results demonstrate comparable performance with other state-of-the-art PET/MR systems. The clinical benefits of high spatial resolution and long axial coverage remain to be further evaluated in specific clinical imaging applications.
Keyphrases
  • pet ct
  • positron emission tomography
  • computed tomography
  • contrast enhanced
  • pet imaging
  • high resolution
  • magnetic resonance
  • magnetic resonance imaging
  • health insurance