An analysis of the spread of electric field within the cochlea for different devices including custom-made electrodes for subtotal cochleoectomy.
Luise WagnerStefan K PlontkeTorsten RahnePublished in: PloS one (2023)
Prediction of speech recognition outcome based on electric field propagation results seems not feasible. Significant differences in field spread between electrode arrays can be clearly demonstrated. This observation and its relevance to hearing treatment and speech recognition should therefore be further investigated in upcoming studies.