Enhancing spinal bone anchor pull-out resistance with an L-shaped anchor.
Esther Paula de KaterMichiel Norbert BlomTeunis Cornelis van DoornQuoc Huy TieuDavid Justin JagerAimée SakesPaul BreedveldPublished in: PloS one (2024)
The success rate of spinal fusion surgery is mainly determined by the fixation strength of the spinal bone anchors. This study explores the use of an L-shaped spinal bone anchor that is intended to establish a macro-shape lock with the posterior cortical layer of the vertebral body, thereby increasing the pull-out resistance of the anchor. The performance of this L-shaped anchor was evaluated in lumbar vertebra phantoms (L1-L5) across four distinct perpendicular orientations (lateral, medial, superior, and inferior). During the pull-out experiments, the pull-out force, and the displacement of the anchor with respect to the vertebra was measured which allowed the determination of the maximal pull-out force (mean: 123 N ± 25 N) and the initial pull-out force, the initial force required to start motion of the anchor (mean: 23 N ± 16 N). Notably, the maximum pull-out force was observed when the anchor engaged the cortical bone layer. The results demonstrate the potential benefits of utilising a spinal bone anchor featuring a macro-shape lock with the cortical bone layer to increase the pull-out force. Combining the macro shape-lock fixation method with the conventional pedicle screw shows the potential to significantly enhance the fixation strength of spinal bone anchors.