Determination of optimum exposure parameters for dentoalveolar structures of the jaws using the CB MercuRay system with cluster signal-to-noise analysis.

Warangkana WeerawanichMayumi ShimizuYohei TakeshitaKazutoshi OkamuraShoko YoshidaGainer R JasaKazunori Yoshiura
Published in: Oral radiology (2018)
No optimum exposure parameters were identified for the evaluation of the lamina dura and periodontal ligament space. D mode was sufficient for the other structures; however, the tube voltage required for each structure differed. Smaller patients required lower tube voltage. I mode, 120 kV may be used for larger lesions.