Optimizing Available Tools for Achieving Result Standardization: Value Added by Joint Committee on Traceability in Laboratory Medicine (JCTLM).
Mauro PanteghiniFederica BragaJohanna E CamaraVincent DelatourKatleen Van UytfangheHubert W VesperTianjiao Zhangnull nullPublished in: Clinical chemistry (2022)
We produced a synopsis of JCTLM-listed higher-order CRMs and RMPs for the selected measurands, including their main characteristics for implementing traceability and fulfilling (or not) the APS for suitable MU. Results showed that traceability to higher-order references can be established by IVD manufacturers within the defined APS for most of the 13 selected measurands. However, some measurands do not yet have suitable CRMs for use as common calibrators. For these measurands, splitting clinical samples with a laboratory performing the RMP may provide a practical alternative for establishing a calibration hierarchy.