Neonatal eyelid penetration from insertion of a fetal scalp electrode: a case report.

Brian T ChengKelly D LaurentiSudhi P Kurup
Published in: BMC pregnancy and childbirth (2022)
Facial or brow presentation during delivery is rare but may increase the risk for misplacement of an FSE. Ultrasound verification of vertex position is warranted immediately prior to placing an FSE for patients at higher risk of facial or brow presentation. Periorbital edema of neonates may protect against damage to deeper structures. However, Ophthalmology should be consulted to rule out ocular injury if the FSE is placed in the periocular region.