Login / Signup

The Gore Cardioform Atrial Septal Defect Occluder: A novel solution to the management of severe hemolysis following transcatheter septal defect closure.

Iman NaimiThomas K JonesZachary Steinberg
Published in: Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions (2022)
Significant hemolysis is a recognized complication of transcatheter high-velocity shunt occlusion using some Amplatzer devices. We describe a case of severe hemolysis following occlusion of an iatrogenic Gerbode defect with an Amplatzer muscular ventricular septal defect occluder successfully managed by transcatheter device removal and reocclusion with a Gore Cardioform Atrial Septal Defect Occluder.