Scintillator-based transverse proton beam profiler for laser-plasma ion sources.

Nicholas P DoverM NishiuchiH SakakiM A AlkhimovaA Ya FaenovY FukudaH KiriyamaA KonK KondoK NishitaniK OguraT A PikuzA S PirozhkovA SagisakaM KandoK Kondo
Published in: The Review of scientific instruments (2018)
A high repetition rate scintillator-based transverse beam profile diagnostic for laser-plasma accelerated proton beams has been designed and commissioned. The proton beam profiler uses differential filtering to provide coarse energy resolution and a flexible design to allow optimisation for expected beam energy range and trade-off between spatial and energy resolution depending on the application. A plastic scintillator detector, imaged with a standard 12-bit scientific camera, allows data to be taken at a high repetition rate. An algorithm encompassing the scintillator non-linearity is described to estimate the proton spectrum at different spatial locations.