Login / Signup

Reliability and performance of the IRRAflow® system for intracranial lavage and evacuation of hematomas-A technical note.

Mette HaldrupMojtaba NazariChenghao GuMads RasmussenStig DyrskogClaus Ziegler SimonsenMads GrønhøjFrantz Rom PoulsenNaveed Ur RehmanAnders Rosendal Korshoej
Published in: PloS one (2024)
Bolting of the IRRAflow® catheter using the Medtronic Silverline® bolt is not recommendable. Increased irrigation rates are recommendable followed by a decrease in drainage bag level. ICP measurement using the IRRAflow® device was unreliable and should be accompanied by a control ICP monitor device in clinical settings.
Keyphrases
  • ultrasound guided