Tibial tray debonding from the cement mantle is associated with deformation of the backside of polyethylene tibial inserts.

Rohan M BhalekarMatthew E NargolNish ShyamAntoni V F NargolStephen R WellsRyan CollierMoreica B PabbruweThomas J JoyceDavid J Langton
Published in: The bone & joint journal (2021)
Explanted TKAs with central locking mechanisms were significantly more likely to debond from the cement mantle. The PE inserts of these designs showed characteristic patterns of deformation, which appeared to relate to the manufacturing process and may be exacerbated in vivo. This pattern of deformation was associated with PE wear occurring at the outer edges of the articulation, potentially increasing the frictional torque generated at this interface. Cite this article: Bone Joint J 2021;103-B(12):1791-1801.