Login / Signup

Unexpected Bare Metal Stent Migration Associated With Aortic Remodeling After PETTICOAT Technique for Aortic Dissection.

Kimihiro KobayashiAtsushi YamashitaYoshinori KurodaShingo NakaiShusuke AraiTetsuro Uchida
Published in: Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists (2022)
Surgeons should consider the potential risk of device migration due to favorable aortic remodeling after TEVAR for aortic dissection, especially with the use of additional bare metal stents.
Keyphrases
  • aortic dissection
  • heart failure
  • human health
  • thoracic surgery
  • risk assessment
  • aortic valve
  • pulmonary arterial hypertension