Login / Signup

Use of Diagnostic Ions for the Detection of Fentanyl Analogs in Human Matrices by LC-QTOF.

Kenneth D SwansonRebecca L ShanerLogan C KrajewskiWilliam A BraggRudolph C JohnsonElizabeth I Hamelin
Published in: Journal of the American Society for Mass Spectrometry (2021)
To combat the ongoing opioid epidemic, our laboratory has developed and evaluated an approach to detect fentanyl analogs in urine and plasma by screening LC-QTOF MS/MS spectra for ions that are diagnostic of the core fentanyl structure. MS/MS data from a training set of 142 fentanyl analogs were used to select the four product ions and six neutral losses that together provided the most complete coverage (97.2%) of the training set compounds. Furthermore, using the diagnostic ion screen against a set of 49 fentanyl analogs not in the training set resulted in 95.9% coverage of those compounds. With this approach, lower reportable limits for fentanyl and a subset of fentanyl-related compounds range from 0.25 to 2.5 ng/mL in urine and 0.5 to 5.0 ng/mL in plasma. This innovative processing method was applied to evaluate simulated exposure samples of remifentanil and carfentanil in water and their metabolites remifentanil acid and norcarfentanil in urine. This flexible approach enables a way to detect emerging fentanyl analogs in clinical samples.
Keyphrases