Attachment of Fibrinogen on Ion Beam Treated Polyurethane.
Vyacheslav S ChudinovIgor N ShardakovIrina KondyurinaAlexey V KondyurinPublished in: Biomimetics (Basel, Switzerland) (2024)
Protein-stable coverage of the artificial implant is a key problem for biocompatibility. In the present study, a protein layer was attached covalently to a polyurethane surface treated by an ion beam. A plasma system consisting of a vacuum chamber (0.8 Pa pressure) with a high voltage electrode powered by a short pulse (20 μS pulse duration and 200 Hz pulse repetition) generator was designed. Polyurethane with a formulation certified as a material for medical implants was treated by nitrogen ions with an energy of 20 keV and 5 × 10 14 -10 16 ions/cm 2 fluence range. Wettability measurements, X-ray photoelectron, Raman, Fourier transform infrared attenuated total reflection, and ellipsometry spectra showed a significant change in the structure of the surface layer of the treated polyurethane. The surface of the treated polyurethane contained a carbonised layer containing condensed aromatic clusters with terminal free radicals. The surface energy of polyurethane surface increased from 33 to 65 mJ/m 2 . The treated polyurethane surface became capable of adsorbing and chemically binding protein (fibrinogen). The designed system for ion beam treatment can be used for surface activation of biomedical polymer devices, where a total protein coverage is required.