Quantitative Assessment of Variation in CT Parameters on Texture Features: Pilot Study Using a Nonanatomic Phantom.
Karen BuchBaojun LiMuhammad M QureshiHirofumi KunoStephan W AndersonOsamu SakaiPublished in: AJNR. American journal of neuroradiology (2017)
Our aim was to evaluate changes in texture features based on variations in CT parameters on a phantom. Scans were performed with varying milliampere, kilovolt, section thickness, pitch, and acquisition mode. Forty-two texture features were extracted by using an in-house-developed Matlab program. Two-tailed t tests and false-detection analyses were performed with significant differences in texture features based on detector array configurations (Q values = 0.001-0.006), section thickness (Q values = 0.0002-0.001), and acquisition mode (Q values = 0.003-0.006). Variations in milliampere and kilovolt had no significant effect.