Accuracy of blood pressure measurement across BMI categories using the OptiBP™ mobile application.

Mary CaillatJean DegottArlene WuerznerMartin ProençainGuillaume BonnierJean-François KnebelChloé StollUrvan ChristenVirginie DurgnatGregory HofmannMichel BurnierGrégoire WuerznerPatrick Schoettker
Published in: Blood pressure (2022)
This study demonstrated that the OptiBP application is easily applicable to overweight and obese participants. Differences between the reference measure and the OptiBP estimation were within ISO limits (criterion 1). In obese participants, the SD of mean error was outside criterion 2 limits. Whether auscultatory measurement, due to arm morphology or the OptiBP is associated with increasing bias in obese still needs to be studied.