Login / Signup

Progress toward a Simplified UTI Diagnostic: Pump-Free Magnetophoresis for E. coli Detection.

Zachary D CallIlhoon JangBrian J GeissDavid S DandyCharles S Henry
Published in: Analytical chemistry (2022)
Urinary tract infections (UTIs) are one of the most common infections across the world and can lead to serious complications such as sepsis if not treated in a timely manner. Uropathogenic Escherichia coli account for 75% of all UTIs. Early diagnosis is crucial to help control UTIs, but current culturing methods are expensive and time-consuming and lack sensitivity. The existing point-of-care methods fall short because they rely on indirect detection from elevated nitrates in urine rather than detecting the actual bacteria causing the infection. Magnetophoresis is a powerful method used to separate and/or isolate cells of interest from complex matrices for analysis. However, magnetophoresis typically requires complex and expensive instrumentation to control flow in microfluidic devices. Coupling magnetophoresis with microfluidic paper-based analytical devices (μPADs) enables pump-free flow control and simple and low-cost operation. Early magnetophoresis μPADs showed detection limits competitive with traditional methods but higher than targets for clinical use. Here, we demonstrate magnetophoresis using hybrid μPADs that rely on capillary action in hydrophilic polyethylene terephthalate channels combined with paper pumps. We were able to detect E. coli with a calculated limit of detection of 2.40 × 10 2 colony-forming units per mL.
Keyphrases