An evaluation of the IOLMaster 700 and its agreement with the IOLMaster v3 in children.

Rebecca E LeightonKaren Mm BreslinKathryn J SaundersSara J McCullough
Published in: Ophthalmic & physiological optics : the journal of the British College of Ophthalmic Opticians (Optometrists) (2021)
When used within a paediatric population, these data demonstrate the IOLMaster 700 to be highly repeatable and reproducible for measures of AL, Kmean , ACD and LT. There is excellent inter-instrument agreement between the IOLMaster 700 and IOLMaster v3 for measures of AL and Kmean . ACD measurements show weaker agreement. These data will be useful when considering reports from population-based studies of refractive error and clinical myopia research.
Keyphrases
  • electronic health record
  • big data
  • emergency department
  • intensive care unit
  • young adults
  • patient reported outcomes
  • deep learning
  • optical coherence tomography