Absence of shock therapy related to improper sensing of noise on the defibrillation test during subcutaneous implantable cardioverter-defibrillator implantation: a case report.
Shota TamuraTomotaka YoshiyamaAtsushi DoiMinoru YoshiyamaPublished in: European heart journal. Case reports (2020)
The present case demonstrated that noise was recorded in two different vectors during DFT, possibly supporting the hypothesis that the muscle spasm of the diaphragm induced by the 50-Hz burst causes oversensing of noise by the S-ICD.