HYPER: pre-clinical device for spatially-confined magnetic particle hyperthermia.

Hayden CarltonMatthias WeberMaximilian PetersNageshwar ArepallyYash Sharad LadAnshul JaswalRobert IvkovAnilchandra AttaluriPatrick Goodwill
Published in: International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group (2023)
We verified the performance of the HYPER to design specifications by independently heating two nanoparticle-rich areas of a phantom placed within the volume occupied by the AMF heating coil.